Friendliness
● Compact design with light weight
● Embedded handle allows for easy transportation
● Straightforward IV set installation
● Intuitive programming procedure with quick guidance on screen
Safety
● High accuracy: ±5%, satisfying neonatal use
● Water-proof membrane on fingers preventing leakage into pump body
● IP34 dust and water protection
Connectivity
● Nurse call/ RS 232/ DC input
Infusion pump
Specifications and functions | |
Dimension |
146 x 193 x 95mm |
Infusion set |
Applicable to GB 8368-2005 disposals; diametersΦ3.5-4.5mm, thickness 0.8-1.2mm |
Infusion mode |
Rate mode |
Flow rate range |
0.1~1500ml/h |
Accumulated volume |
0.1~9999 ml |
VTBI |
0.1~999.9 ml, increment is 0.1 ml; 1000-9999ml, increment is 1ml |
Increment |
0.1~999.9 ml/h, increment is 0.1 ml/h; 1000~1500 ml/h, increment is 1 ml/h |
Accuracy |
≤ ±5% |
KVO rate |
0.5 ml/h |
Bolus rate |
0.10-1500ml/h |
Purge rate |
600ml/h |
Occlusion |
Low 150±125 mmHg (20kPa); Medium 525±125 mmHg (70kPa); High 900±180 mmHg (120 kPa) |
Air bubble detection |
Single bubble detection: 50, 100, 250, 500, 800l μl |
Screen |
2.5" LCD monochrome screen, 132 x 32 pixels |
Titration |
Yes, increment is 0.1ml/h |
Keypad lock |
Yes |
Pause |
Yes |
Self-test system |
Yes |
Weight |
<1.6kg (not including clamp) |
Handle |
Embedded handle |
Mounting type |
Standard: Vertical Pole |
Alarm Volume |
1-8 levels selectable |
Language |
8, Chinese, English, Spanish, French, Russian, Turkish, Portuguese, German |
Alarm | |
VTBI done, VTBI near done, KVO finish, Bottle empty, door open, IV set not installed, occlusion, air in line, drop error, system error, system abnormal, battery low, battery empty, reminder, AC power disconnection, standby end |
|
Safety Requirement | |
AC power |
100-240V (50/60Hz, 0.4-0.14A) |
DC power |
10-16V (2.25-1.5A) |
Battery type |
Lithium |
Battery operation time |
>4 hours@25ml/h |
Charging time |
≤6 hours@100% |
Classification |
CF, Class I |
IP Grade |
IP 34 |
Temperature |
5~40ºC for operating, -20~60ºC for storage |
Humidity |
15~95% for operating, 10~95% for storage |
Air pressure |
57~106kPa for operating, 50~106kPa for storage |
Interface | |
Data interface |
Supported, RS232 |
Nurse Call |
Supported |
DC input |
Supported |
http://www.vetmeditech.com/